Erich Wiesmueller Analysegeraete/ Spektrometer
Tel.: +43 (0) 2262 72027 E-Mail: office.wiesmueller@gmail.com

english Deutsch

  • science lab equipment
  • lab equipment
  • Analyzers spectrometry
  • Quality Control Spectrometers

nitrogen/total sulfur combustion analyzer | XPLORER-NS & XPLORER-TX/TS

More Views

Attachments

nitrogen/total sulfur combustion analyzer | XPLORER-NS & XPLORER-TX/TS
€1,000,000.00

Overview

The XPLORER-NS was developed as a total nitrogen/total sulfur combustion analyzer, and the XPLORER-TX/TS a total halogen and sulfur microcoulometric combusiton analyzer.  Both instruments offer fast, precise analysis of all phases; liquids, solids, and gases (including LPG's). These instruments were designed to offer standardized and customized solutions to match both current and future analytical needs, ranging from low ppb’s to high ppm’s.

Manufactured by TE Instruments.

Features

  • Ultra-low detection limits, high stability and reliability due to the temperature controlled detectors and feedback control loop.
  • Compact design, with the smallest footprint in the market.
  • Short start-up time. (Less than 15 minutes from standby mode).
  • Fast and precise measurement of liquids, solids, and gases (including LPGs).
  • Automation available for each type of measurement.
  • Intuitive user interface and operational software.
  • Simultaneous analysis of halogens and sulfur.
  • Fast and easy switching between modules, resulting in high productivity.
  • Complies with international standards like: ASTM, ISO, EN and IP.
  • Fully automated creation of calibration curves from a single stock solution with the optional ARCHIE autosampler.
  • Fast generation of sample lists and application methods.
  • Low maintenance due to optimized combustion and the proper conditioning of the resulting gases produces near zero downtime.

Specifications

XPLORER-TX/TS System Specifications

Dimensions 

(w x h x d) 36 x 27.2 x 69 cm (14.2 x 10.7 x 27.2 in.)

Weight 

27 kg (59.5 lbs) without furnace tube and introduction module

Voltage 

100-240 V, 50-60 hz

Power requirement (maximum)

 1.15KW

Gas Connectors 

⅛” Swagelok

Gases required 

Oxygen 99.6% (2.6), Argon 99.998% (4.8) or Helium 99.99%

Input gas pressure 

3-10 bar (45-145 PSI)

Internal gas pressure 

1.8 bar, adjustable

Furnace voltage 

Dual zone, low voltage

Furnace temperature (maximum) 

1150 °C (2102 °F)

Furnace cooling 

Pulling fan with auto control

Sample introduction 

Quartz boat

Solids 

5-1000 mg

Boat driver 

Software controlled, adjustable

Slider/shutter driver 

Software controlled, adjustable

Detector: 

SMD, Digital coulometer

Detector accuracy: 

Better than 2% CV

Titration cell conditioning: 

Temperature controlled, adjustable

Software 

dot.NET-based

Ambient temperature 

5-35 °C (41-95 °F) non-condensing

XPLORER-NS System Specifications

Dimensions 

(w x h x d) 36 x 27.2 x 69 cm (14.2 x 10.7 x 27.2 in.)

Weight 

32 kg (70.5 lbs) without furnace tube and introduction module

Voltage 

100-240 V, 50-60 hz

Power requirement (maximum)

 1.15KW

Gas Connectors 

⅛” Swagelok

Gases required 

Oxygen 99.6% (2.6), Argon 99.998% (4.8) or Helium 99.99%

Input gas pressure 

3-10 bar (45-145 PSI)

Internal gas pressure 

1.8 bar, adjustable

Furnace voltage 

Dual zone, low voltage

Furnace temperature (maximum) 

1150 °C (2102 °F)

Furnace cooling 

Pulling fan with auto control

Sample introduction 

Solid or liquid by boat, Liquid by direct injection, Gases and LPGs by GLS

Sample size

Solids:5-1000 mg; Liquids: 100 μL; Gas; 10 ml; LPG; 100 μL

Semi-automatic boat/syringe driver

Software controlled, programmable method fi le

Slider/shutter driver 

Software controlled, adjustable

Detector Nitrogen

Chemiluminescence

Detector Sulfur

Xenon pulsed UV-fl uorescence AFC Technology

Detector accuracy

Temperature controlled, adjustable

Software 

Better than 2% CV

Detector conditioning

Temperature controlled, adjustable

Vacuum pump

Internal 24 volt DC

Software

dot.NET-based, TEIS software

Ambient temperature

5-35 °C (41-95 °F) non condensing

Availability: In stock

OR
Description
Additional Information
Reviews

Product Tags

Use spaces to separate tags. Use single quotes (') for phrases.

Used Spark Spectrometers | Spark Spectrometer | Optical Emission Spectrometers | X Ray Diffractometer